G005858
LC/MS Analysis of 25-Hydroxyvitamin D2 and 25-Hydroxyvitamin D3 on Ascentis® Express F5
application for HPLC
About This Item
Recommended Products
technique(s)
HPLC: suitable
test parameters
column: Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)
mobile phase: 5 mM ammonium formate in methanol:water (75:25)
flow rate: 0.4 mL/min
column temp.: 40 °C
detector: UV, 265 nm (MS, ESI(+), m/z 100-1000)
injection: 1 μL
sample: 20 μg/mL in methanol
suitability
application for HPLC
application(s)
clinical
vitamins, nutraceuticals, and natural products
Analysis Note
Legal Information
Analyte
analytical column
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service